How can a standard adjust be set on the DR900?

Document ID

Document ID TE8483

Published Date

Published Date 12/20/2021
Print
Question
How can a standard adjust be set on the DR900?
Summary
DR900 standard adjust
Answer
To set a standard adjust on the DR900 Multiparameter Portable Colorimeter (Product # 9385100), first select the program needed to adjust.
Once the program is set, run a standard of known concentration for the specific test.
When the results are on the display, select Options --> Advanced Options --> Standard Adjust.
On this screen, you will set the value to which the standard will be adjusted.
The value can be adjusted up to 10% of the displayed readout.