What are the instructions for a Chemkey calibration (standard adjust) on the SL1000 Portable Parallel Analyzer?

Document ID

Document ID TE8293

Published Date

Published Date 10/22/2019
Print
Question
What are the instructions for a Chemkey calibration (standard adjust) on the SL1000 Portable Parallel Analyzer?
Summary
Standard adjust instructions on the SL1000
Answer
To be able to perform Chemkey Calibration (standard adjust) on the SL1000 Portable Parallel Analyzer (PPA), one must first change the calibration option settings from factory to standard adjust and then perform a standard adjust calibration. To do this, follow these steps:

1. Insert the chemkey to be adjusted into any of the available 4 slots
2. Select the 'Main Menu' button
3. Select 'Settings' 
4. Select 'Chemkey'
5. Select the applicable chemkey that will be adjusted
6. Select 'Calibration Method'
7. Change the method from 'Factory' to Standard Adjust' (or vice versa if one is wanting to go back to the factory settings)
8. Click 'OK' to save these changes and press the 'Exit' button 3 times to get back to the 'Main Menu'
9. Select 'Calibrate' under the main menu settings
10. Select the applicable chemkey to be adjusted
11. Enter the standard adjust value and press 'OK'
12. Complete the instructions shown on the display to calibrate the chemkey (be sure to have a made up primary standard ready to be measured to complete this process)

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.