Can UVAS be used to correlate a Sulfide measurement?

Document ID

Document ID TE8265

Version

Version 1.0

Status

Status Published

Published Date

Published Date 08/02/2017
Question
Can UVAS be used to correlate a Sulfide measurement?
Summary
It is not possible to correlate a sulfide measurement by using UVAS.
Answer
Sulfide is an inorganic component. It is not possible to correlate a sulfide measurement by using UVAS. You may be able to determine a correlation to parameters like COD, TOC, DOC, BOD, e.g. dissolved organic components that absorb UV light at 254nm wavelength.  UVAS measures UV absorption, it is not a COD, TOC or BOD sensor. If TOC results are required for complying with a regulation or NPDES permit, an online TOC analyzer is the only solution.

 