Question
Why does the Flo-Tote 3 sensor show excessively high level?
Summary
Excessively high level readings with the Flo-Tote 3.
Answer
When the Flo-Tote 3 is showing level in excess of 100-300 inches (254-762 cm), the level sensor has failed and the sensor needs to be replaced. Please note that the maximum calibration factor for the Flo-Tote 3 is 12 inches (30.48 cm).