How do you program a velocity only (wafer) sensor on a Sigma 950?

Document ID

Document ID TE8133

Version

Version 1.0

Status

Status Published

Published Date

Published Date 07/24/2017
Question
How do you program a velocity only (wafer) sensor on a Sigma 950?
Summary
Programming the velocity only (wafer) sensor on a Sigma 950.
Answer
  1. From the Main Menu select Setup, modify Selected Items.  
  2. Select the Velocity Direction by using the up and down keys, then press Select.
  3. Select the Velocity Direction (upstream, downstream, always positive) by using the Change Choice key, then press Accept.
  4. Highlight the velocity units by using the up and down keys, then press Select to continue.
  5. Select the velocity units by using the up and down keys, then press Accept to continue.
  6. Highlight the Velocity Cutoff by using the up and down keys, then press Select to continue.
  7. Read the velocity cutoff information screen. Press any key to continue.
  8. Set the Velocity Cut Off using the keypad. Press Accept to continue
  9. Set the Velocity Default, by using the keypad. Then press Accept.  