How do you program a ORP probe on a Sigma 950?

Document ID

Document ID TE8123

Version

Version 1.0

Status

Status Published

Published Date

Published Date 07/24/2017
Question
How do you program a ORP probe on a Sigma 950?
Summary
Programming the ORP probe on a Sigma 950.
Answer
  1. From the Main Menu select Options, Advanced Options, Datalog. 
  2. Select the inputs by using the up and down keys, then press Select.
  3. Select the ORP probe by using the up and down keys, then press Select.
  4. Press Change Choice to cycle between logged and not logged, then press Accept for when your choice is highlighted.
  5. Enter a logging interval using the numerical keypad, then press Accept. Note: all valid intervals are in the status bar at the bottom of the screen.
  6. The meter is now logging the ORP probe. Press Return and then Main Menu to get back to the main menu. 