Question
For the Glass Rod Secondary Turbidity Standard what should the acceptance range bet set to for the TU5200?
Summary
Acceptance range of Glass Rod Secondary Turbidity Standard <0.1 NTU for TU5200, TU5300 sc, and TU5400 sc Laser Turbidimeter {Product # LZY901)
Answer
To set up the TU5200 to use the Glass Rod Secondary Turbidity Standard <0.1 NTU for TU5200, TU5300 sc, and TU5400 sc Laser Turbidimeter {Product # LZY901), do the following steps.
1. Clean the vial with a no-lint cloth (LZY945) to remove contamination. Keep contamination off the glass. Refer to Prevent vial contamination in the TU5200 user manual for additional tips on keeping the glass clean.
2. On the TU5200, Push Verification.
3. Push Options > Verification Setup.
4. Push Acceptance Range
5. Change the acceptance range to 0.015 NTU.
Please see attached documentation in English, French, Spanish, German, Italian, Chinese.
1. Clean the vial with a no-lint cloth (LZY945) to remove contamination. Keep contamination off the glass. Refer to Prevent vial contamination in the TU5200 user manual for additional tips on keeping the glass clean.
2. On the TU5200, Push Verification.
3. Push Options > Verification Setup.
4. Push Acceptance Range
5. Change the acceptance range to 0.015 NTU.
Please see attached documentation in English, French, Spanish, German, Italian, Chinese.