What could cause very small peaks when running samples and standards on a Lachat Flow Injection Analysis Method?

Document ID

Document ID TE7989

Version

Version 3.0

Status

Status Published

Published Date

Published Date 05/21/2019
Question
What could cause very small peaks when running samples and standards on a Lachat Flow Injection Analysis Method?
Summary
Small peaks when running Lachat Flow Injection Analysis (FIA) methods.
Answer
If peaks are significantly smaller than the support data on a Lachat FIA method, there are several things that could be at fault:
  • Verify the valve timing. Run dye as a sample and verify the sample loop is being filled properly and injected into the manifold.
  • There could be a valve issue. Try using a different injection valve.
  • Verify the reagents are fresh and free from contamination. Old source chemicals can also be a concern even if the reagent was freshly prepared.
  • Verify the pH of the reagents, samples and standards against the method.
  • Check the standard source and preparation.
  • Verify the correct sample loop for the method is used.
  • Verify the correct pump tubes are being used. Check both the size and the material.
  • Verify the correct detector filter is used.
  • Try using a different detector to verify the detector is functioning properly.
  • Check the baseline voltage.
  • Check the pump to verify it is working at the correct speed.
How is valve timing optimized on the Lachat QC8500 FIA?
 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.