What is the minimum immersion depth for the Intellical™ CDC401 Laboratory 4-Poles Graphite Conductivity Cell?

Document ID

Document ID TE7982

Published Date

Published Date 07/26/2018
Print
Question
What is the minimum immersion depth for the Intellical™ CDC401 Laboratory 4-Poles Graphite Conductivity Cell?
Summary
Minimum sample volume for lab conductivity probe.
Answer
The probe has a diameter of 16 mm and a minimum immersion depth of 45 mm. So it would require that the volume level of the sample in the container when the measurement is taken with the volume displaced by the probe once the probe is submerged to be 45 mm up the side of the probe which is enough to completely submerge the window of the probe.

The temperature sensor and poles of the electrode are all inside the window of the probe. So there it is not problematic if outside walls or the bottom of the probe come in contact with the walls of the container that the measurement is taken in.