How do you get the data of the current calibration on a Sension+ PH1 Portable pH meter?

Document ID

Document ID TE7938

Version

Version 2.0

Status

Status Published

Published Date

Published Date 06/12/2018
Question
How do you get the data of the current calibration on a Sension+ PH1 Portable pH meter?
Summary
Accessing calibration data on a Sension+ PH1 Portable pH meter
Answer
Follow these instructions for viewing the slope of the current calibration on the Sension+ PH1 Portable pH Meter (Product # LPV2500.97.0002).
1. From the home screen, press the Cal button To enter the Calibration menu.

2. From here you can press the down arrow to change it to ORP. Otherwise if you're pulling up the pH calibration data, skip to step 3.

3. If the parameter in the center of the display is correct. Press the Cal button  to accept and continue to the next screen. For pH this adds the CAL1 to the bottom of the screen.

4. In this screen press the up arrow and down arrow at the same time. This will take you to the screens that display the calibration data for the current calibration.
For pH

For ORP

 