Why is LDO101 greyed out on the SL1000 when trying to calibrate?

Document ID

Document ID TE7921

Version

Version 2.0

Status

Status Published

Published Date

Published Date 08/28/2018
Question
Why is LDO101 greyed out on the SL1000 when trying to calibrate?
Summary
SL1000 and LDO101 calibration.
Answer
With the SL1000 Portable Parallel Analyzer™ (PPA); (Product # 9430000) follow these steps:

1. Push Main Menu > Settings > Probe
2. Select LDO Options > Calibration Options > Calibration Method > User Cal - 100%

This will allow a water saturated air calibration and Intellical™ LDO101 Field Luminescent/Optical Dissolved Oxygen (DO) Sensor to be selected and not be greyed out.