Why is LDO101 greyed out on the SL1000 when trying to calibrate?

Document ID

Document ID TE7921

Published Date

Published Date 10/22/2019
Print
Question
Why is LDO101 greyed out on the SL1000 when trying to calibrate?
Summary
SL1000 and LDO101 calibration.
Answer
With the SL1000 Portable Parallel Analyzer™ (PPA); (Product # 9430000) follow these steps:

1. Push Main Menu > Settings > Probe
2. Select LDO Options > Calibration Options > Calibration Method > User Cal - 100%

This will allow a water saturated air calibration and Intellical™ LDO101 Field Luminescent/Optical Dissolved Oxygen (DO) Sensor to be selected and not be greyed out.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.