What data is stored in the reading log for the SL1000?

Document ID

Document ID TE7909

Published Date

Published Date 08/28/2018
Print
Question
What data is stored in the reading log for the SL1000?
Summary
Reading log data.
Answer
For ChemKey results the time and date, Parameter, Lot info for the ChemKey used, what slot the ChemKey was in when the measurement was taken, the chemical form, the mg/L result, the absorbance measurement (ABS), and percent transmittance measurement (%T) are all stored and viewable in the reading log.

For pH readings the time and date, model number of the probe used, serial number of probe used, pH result, and temperature result are all stored and viewable in the reading log.

For Conductivity the time and date, model number of the probe used, serial number of probe used, Conductivity result (µS/cm), temperature result, and TDS result (mg/L) are all stored and viewable in the reading log.

For Dissolved Oxygen the time and date, model number of the probe used, serial number of probe used, percent saturation result, temperature result, pressure result, and mg/L result are all stored and viewable in the reading log.

For ISE readings  the time and date, model number of the probe used, serial number of probe used, mg/L result, temperature result, and mV result are all stored and viewable in the reading log.

For the Chlorine Verification ChemKey the time and date, Chlorine, lot of the verification ChemKey, slot checked, chemical form, mg/L result, abs result, %T result, expected result, and pass or fail notification are all stored and viewable in the reading log.

For the System Verification ChemKey, Verification, lot of the verification ChemKey, slot checked, chemical form (Verification), abs results, %T result, expected result, and pass or fail notification are all stored and viewable in the reading log.