What is the minimum required pulse width (mSec) required on the AS950 sampler?

Document ID

Document ID TE7895

Version

Version 1.0

Status

Status Published

Published Date

Published Date 06/21/2017
Question
What is the minimum required pulse width (mSec) required on the AS950 sampler?
Summary
Minimum pulse width required for AS950
Answer
The AS950 documentation states that a minimum pulse of 50 mSec is required. However, it has been determined that there is a 20 mSec lag (or bounce time) in the processor. So the minimum pulse required is actually 70 mSec.