Should the UVAS sc probe be installed upstream or downstream of a UV disinfection system?

Document ID

Document ID TE7865

Version

Version 1.0

Status

Status Published

Published Date

Published Date 06/19/2017
Question
Should the UVAS sc probe be installed upstream or downstream of a UV disinfection system?
Summary
UVAS sc probe and installation location near UV disinfection system
Answer
It does not matter where the UVAS sc probe is installed in relation to the UV disinfection system.  The UVAS sc probe is monitoring the organic load in the water.  The UV disinfection system is altering the DNA of the organisms in the water, but does not alter the organic load.