Question
Can the Refractometer, Automotive differentiate between Ethylene glycol and Propylene glycol?
Summary
Ethylene glycol and Propylene glycol on Refractometer, Automotive
 
Answer
The Refractometer, Automotive (Product # 2300055) does not differentiate between Ethylene glycol and Propylene glycol. There are two separate functions available for each parameter, however, the results determined are based off the refractive index of the solution compared against distilled water. If, for example, the solution contained both Ethylene glycol and Propylene glycol, each substance could not be analyzed independently.
 