Can users create custom calibration curves in the TL23 series turbidimeters?

Document ID

Document ID TE7805

Published Date

Published Date 06/16/2017
Print
Question
Can users create custom calibration curves in the TL23 series turbidimeters?
Summary
Accessories needed for customized calibration.
Answer
Yes, users can create up to 5 custom calibrations with a maximum of 5 standards plus the blank; those calibrations are done in standard vials.

11 mm vials require a special user program. Contact technical support to further discuss this application.

As custom calibrations may not be linear, the TL23 series turbidimeters offer different options to fit the curve.  This is different from standard calibration curves which can’t be shown as the span for standard calibration is too large.
 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.