What is the recommended seed contribution for BOD?

Document ID

Document ID TE7788

Version

Version 4.0

Status

Status Published

Published Date

Published Date 10/09/2018
Question
What is the recommended seed contribution for BOD?
Summary
Seed contribution for BOD
Answer
When testing Biochemical Oxygen Demand (BOD), Standard Methods recommends that the seed contribution be 0.6 - 1.0 mg/L oxygen depletion, though this can be adjusted to ensure Glucose and Glutamic Acid (GGA) standards read 198 +/- 30.5 mg/L.

Additional troubleshooting info can be found in the following article:

What should be done if the BOD Seed Contribution is too high or too low?


 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.