What is the recommended seed contribution for BOD?

Document ID

Document ID TE7788

Published Date

Published Date 10/09/2019
Question
What is the recommended seed contribution for BOD?
Summary
Seed contribution for BOD
Answer
When testing Biochemical Oxygen Demand (BOD), Standard Methods recommends that the seed contribution be 0.6 - 1.0 mg/L oxygen depletion, though this can be adjusted to ensure Glucose and Glutamic Acid (GGA) standards read 198 +/- 30.5 mg/L.

Additional troubleshooting info can be found in the following article:

What should be done if the BOD Seed Contribution is too high or too low?


 