Is there a way to trigger a relay if there is no flow to a CL10 chlorine analyzer?

Document ID

Document ID TE7697

Version

Version 1.0

Status

Status Published

Published Date

Published Date 06/06/2017
Question
Is there a way to trigger a relay if there is no flow to a CL10 chlorine analyzer?
Summary
Relay trigger for no flow on CL10 analyzer
Answer
To trigger a relay for no flow on CL10:

sc-controller menu
sc200 setup >  relay setup > (choose a relay) > select source > CL2 gateway > set function > warning > activation > warning level (no flow).