Why does DPD continue to develop color over time?

Document ID

Document ID TE7608

Published Date

Published Date 09/08/2020
Print
Question
Why does DPD continue to develop color over time?
Summary
DPD chlorine reagent.
Answer
DPD (N,N-diethyl-p-phenylenediamine) will continue to oxidize due to contact with the atmosphere over time and it will eventually turn a very dark brown. This is why timing is important in Free and Total Chlorine tests. Free Chlorine must be measured within a minute of adding the reagent. Total chlorine must be measured within three to six minutes after adding the reagent.