Why does DPD continue to develop color over time?

Document ID

Document ID TE7608

Published Date

Published Date 02/13/2020
Print
Question
Why does DPD continue to develop color over time?
Summary
DPD chlorine reagent.
Answer
DPD (N,N-diethyl-p-phenylenediamine) will continue to oxidize due to contact with the atmosphere over time and it will eventually turn a very dark brown. This is why timing is important in Free and Total Chlorine tests. Free Chlorine must be measured within a minute of adding the reagent. Total chlorine must be measured within three to six minutes after adding the reagent.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.