What is the air purge system on the TL23 Series Benchtop Turbidimeters?

Document ID

Document ID TE7587

Version

Version 4.0

Status

Status Published

Published Date

Published Date 09/21/2018
Question
What is the air purge system on the TL23 Series Benchtop Turbidimeters?
Summary
Air purge system on TL23 Series Benchtop Turbidimeters.
Answer
The air purge system on the TL23 Series Benchtop Turbidimeters is used to keep condensation off the external surface of the sample cells when cold samples are measured.

The air purge system pushes dry air through the optical compartment to keep the outside of the sample cell dry. The connection is made at the air purge fitting on the back of the instrument.

Hach does not sell an air purge system, but provides a fitting on the back of the instrument where one could be connected. The air purge system would need to be purchased from another company.

More information, as well as a recommendation for a Balston manufactured air purge system, is given in the user manuals for the respective TL2300, TL2310, TL2350, and TL2360 turbidimeters.