What is the proper procedure for calibrating the U53 ultrasonic sensor?

Document ID

Document ID TE7580

Version

Version 2.0

Status

Status Published

Published Date

Published Date 10/25/2018
Question
What is the proper procedure for calibrating the U53 ultrasonic sensor?
Summary
One Point Level Calibration on the U53 ultrasonic sensor.
Answer
  1. Menu > Sensor Setup > Calibrate
  2. Select 1 pt Depth Cal, ENTER
  3. Select Active, ENTER
  4. Wait for sensor depth to stabilize, ENTER
  5. Measure the depth of the water under the sensor and enter value (set Depth), ENTER
  6. Will Display Passed or Failed & Select if a new sensor or not
  7. Calibration complete