What is the proper procedure for resetting/adjusting the totalizer on the sc200 controller?

Document ID

Document ID TE7577

Published Date

Published Date 12/20/2021
Question
What is the proper procedure for resetting/adjusting the totalizer on the sc200 controller?
Summary
Procedure for resetting/adjusting when 'Totalizer is Full +++++++++' displaying on screen
Answer

If the totalizer is set in Auto mode, once the totalizer is full, it will automatically reset. If the volume multiplier is set to Auto it will change from x1, x10, x100 and so on until the max value is reached and then it will Auto reset.

If the totalizer is set in Manual mode, it will require user to select the Reset Totalizer option and manually reset the totalizer.