What is the proper procedure for programming the Sigma 950 for sample pacing?

Document ID

Document ID TE7540

Version

Version 1.0

Status

Status Published

Published Date

Published Date 07/11/2017
Question
What is the proper procedure for programming the Sigma 950 for sample pacing?
Summary
Programming the Sigma 950 for sample pacing.
Answer
  1. From the Main Menu screen, select Setup, Modify Selected Items, Sample Pacing, then click Select
  2. The meter will show you if Sample Pacing is Enabled or Disabled. Select change choice to switch between the two settings, then select Accept.  
  3. Select Clear Entry, then enter in a Trigger Value using the keypad. Also select the units for the trigger value.
  4. To change the units for the trigger value, select the Change Units key until the correct units show up next to the trigger value.
  5. Once you have selected the correct trigger value and units, select Accept.
  6. The flow meter is now programmed to trigger the sampler every time the trigger value is met.