What is the proper procedure to restore the FL900 to its factory defaults using FSData?

Document ID

Document ID TE7530

Version

Version 1.0

Status

Status Published

Published Date

Published Date 07/12/2017
Question
What is the proper procedure to restore the FL900 to its factory defaults using FSData?
Summary
Restoring the FL900 to its factory defaults using FSData.
Answer
  1. Connect to the Logger.
  2. On the top left, there is a Hach Logo and then "FSData Desktop".  To the right of that is a small triangle.  Click on it.
  3. A box will come up.  On the left near the bottom is "Instrument Tools."  Click on it.
  4. This will take you to a page that has 3 Buttons.  Click on the button that says "Restore."
  5. A box will come up warning you that proceeding will erase all data, logs and the program from the FL900.  If you want any of these, cancel and save them.
  6. Wait.  It will take around 2 minutes.
  7. A box will come up that says "The factory defaults were successfully restored in the connected FL900."
  8. You have now restored the FL900 to its factory defaults.