Why won’t FSData Desktop communicate with the FL900?

Document ID

Document ID TE7519

Version

Version 2.0

Status

Status Published

Published Date

Published Date 03/13/2018
Question
Why won’t FSData Desktop communicate with the FL900?
Summary
Troubleshooting FL900 to computer (with FSData Desktop) communication problems.
Answer
  1. Verify the FL900 has power.  Check the batteries, AC power or solar panel power.
  2. If you are using the USB port on your computer, remove the USB connector from the computer and re-insert it into the USB port.  Try another USB port.
  3. Reboot the logger:   Remove power.  Wait 5 seconds.  Reconnect power to the logger (the reboot takes 3-4 minutes).
  4. If your cable is an FL900 to laptop cable with an RS232 and a “Serial to USB converter,” you may need a proper Serial to USB converter as all types do not work.  We recommend the Sewell SW-1314 (Hach p/n 6249200, or you can order it directly from Sewell).
  5. If you have an RS232 cable, verify the baud rate is the same on the FL900 and FSData Desktop.
  6. The program could be corrupt.  Reinstall the FSData Desktop program.  See What software is used to access an FL900 flow sensor?
  7. The logger may be bad.  If so, it will need to be sent back for repair.
  8. If you cannot communicate with the FL900, call the Hach Flow Tech Support Line at 800-368-2723 ext 6063 or email us at FlowTechSupport@hach.com.