What is the proper procedure to set/remove a passcode on the sc200?

Document ID

Document ID TE7502

Version

Version 1.0

Status

Status Published

Published Date

Published Date 06/23/2017
Question
What is the proper procedure to set/remove a passcode on the sc200?
Summary
sc200 Passcode
Answer
Enable/Disable
  1. Menu > sc200 Settings > Enter
  2. Security Setup > Enter
  3. Set Passcode > Enter
  4. Select Disabled or Enabled > Enter
 
Edit Passcode
  1. Only available if passcode is enabled
  2. Menu > sc200 Settings > Enter
  3. Security Setup > Enter
  4. Set Passcode > Enter
  5. Use the arrow keys to enter the current valid passcode > Enter
  6. Select Edit Pass Code > Enter
  7. Push the HOME key or perform a controller restart.
  8. The new passcode settings are saved, and the new passcode is required to enter the Security Setup, Datalog Setup and Test/Maint menus.