How can I validate my 5500 Ammonia/ Monochloramine (AMC) analyzer?

Document ID

Document ID TE7429

Published Date

Published Date 07/14/2020
Print
Question
How can I validate my 5500 Ammonia/ Monochloramine (AMC) analyzer?
Summary
5500 AMC analyzer validation testing.
Answer
To validate the 5500 AMC analyzer, take 150 mL of each of the two standards and mix them together. You will then run this 1 ppm AMC standard on both an SL1000 using the 9429600 Chemkey and in the Grab Sample mode on the 5500 AMC analyzer using 150 mL for each unit.