Can the DR 1900 read standards directly to create a user entered program calibration curve?
Summary
DR 1900 user program absorbance
Answer
The DR 1900 cannot read standards to create a calibration curve within the User Program creation. 
To obtain the absorbance values for each calibration standard to enter in for a calibration curve, it is recommended to read and record each standard using the single wavelength feature of the spectrophotometer at the appropriate wavelength first, and then copy those values into the 'Enter Values' portion of the user program creation. 