How are the batteries changed on the DR1900?

Document ID

Document ID TE7368

Version

Version 4.0

Status

Status Published

Published Date

Published Date 10/02/2018
Question
How are the batteries changed on the DR1900?
Summary
DR1900 battery change
Answer
If the DR1900 Portable Spectrophotometer Product # DR1900-01H will not turn on due to the batteries having no charge, unscrew the plate from the back of the instrument using a Phillips head screwdriver and replace the batteries.
The instrument uses four AA batteries.