Question
Is there an inline mounting option for the Solitax sc sensor that is not a ball valve?
Summary
Inline mounting option for Solitax sc sensor.
Answer
Catalog Number/ Order Code LZX461 (Installation armature, inline fitting, stainless steel, for Solitax inline sc and Solitax highline sc) is an inline mounting option for the Solitax sc sensor that does not have a ball valve for removal.  In order to use this option, there should be a service shut off upstream of the sensor, so the pipe can be depressurized and drained down before removing the sensor for maintenance.  Please see the attached mounting hardware instructions for more information.