When does the 5500sc analyzer compressor turn on and off?

Document ID

Document ID TE7350

Published Date

Published Date 01/06/2022
Print
Question
When does the 5500sc analyzer compressor turn on and off?
Summary
5500sc analyzer compressor operation.
 
Answer
PRESSURE REGULATION ON 5500sc ANALYZER:
  • The reagent air pressure is regulated at 4.00 psi (3.95 to 4.10 psi).
  • The air compressor will not come on until the pressure drops below 3.95 psi.
  • The compressor will turn off when greater than 3.95 psi but there will be an overshoot so there is a delay of ~1 second to let the pressure stabilize before deciding to turn the compressor on again or turn the pressure relief valve on if over 4.10 psi or 3; if the pressure is between 3.95 to 4.10 no changes occur. 
  • The air compressor or relief valve (which has an air restriction to slow the air flow down) does not operate when delivering reagents so that the pressure is stable, but when the standards are flowing the pump can come on due to the larger volume of solution that's delivered.
  • Even if the pressure is somewhere between 3.95 to 4.10 psi, the current measured psi value is used in the equation to calculate the reagent delivery time.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.