How should the sample blank be prepared for the Nickel TNTplus Vial Test (0.1-6.0 mg/L Ni)?

Document ID

Document ID TE7247

Version

Version 4.0

Status

Status Published

Published Date

Published Date 04/16/2019
Question
How should the sample blank be prepared for the Nickel TNTplus Vial Test (0.1-6.0 mg/L Ni)?
Summary
Sample blank for Nickel TNT856
Answer
To determine the sample blank for Nickel TNTplus Vial Test (0.1-6.0 mg/L Ni) (Product # TNT856), run the procedure as written in method 10220, but substitute 0.2 mL of deionized water in place of the 0.2 mL of Solution A in step 3. Use the red stopper to cap the sample vial. The value obtained in step 6 is then subtracted from the value obtained on the original sample to give the corrected sample concentration.

Alternatively, samples that contain only turbidity may be first filtered through a membrane filter and then analyzed. Report the results as soluble nickel.

Samples without color or turbidity do not require sample blanks.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.