What does the MEMORY FULL message on the TSS portable mean and how is the message resolved?

Document ID

Document ID TE7200

Version

Version 5.0

Status

Status Published

Published Date

Published Date 10/02/2018
Question
What does the MEMORY FULL message on the TSS portable mean and how is the message resolved?
Summary
TSS Portable Memory Full
Answer
The TSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Level System stores measurement results in it's internal memory, up to 290 points including calibration points (see also: Does the specification of 290 measurement values only include what is stored in the DATA menu?). If the memory is full it will display the MEMORY FULL message when an additional measurement is started and will not allow another measurement to start until the data log is deleted.

To do this select DATA from the home screen and confirm with ENTER/ON, then select DELETE MEP and confirm with ENTER/ON. Select YES and confirm with ENTER/ON.
This will only delete the measurement points that were obtained with the selected curve and none of the calibration points. Measurement points stored on other curves will not be deleted.
 