What is the difference between the DR3900 and DR6000?

Document ID

Document ID TE5743

Published Date

Published Date 12/20/2021
Question
What is the difference between the DR3900 and DR6000?
Summary
DR3900 vs. DR6000
Answer

The biggest different between the DR3900 and DR6000 is that the DR6000 has a UV lamp.

The DR6000 has an automatic moving carousel.

There are certain software applications that can only be used on the DR6000. These include the Brewery Analysis Package (Catalog Number/ Order Code LZV942), the Drinking Water Analysis Package (Catalog Number/ Order Code LZV925), and the Enzymatic Food Analysis Software (Catalog Number/ Order Code LZV943).

The DR6000 has tighter specifications than the DR3900. For example, the wavelength accuracy of the DR6000 is ± 1 nm. The wavelength accuracy of the DR3900 is ± 1.5 nm.
The data sheets for the DR3900 and DR6000 can both be found on the website.
 