Question
What is the procedure for replacing the filling solution in a refillable pH probe?
Summary
Instructions for replacing pH probe filling solution.
Answer
-
Remove old/contaminated filling solution in the probe by tilting the solution back towards the filling hole, a pipet such as Pipet, Sterile, Transfer, 15/pk (Catalog Number/Order Code 2232512) may be used to remove the solution.
-
Using a transfer pipette, add some deionized (DI) water, such as Water, Deionized, 100 mL (Catalog Number/Order Code 27242) to the probe through the filling hole.
-
Shake the probe to rinse off the inside of the probe with the DI water.
-
Use a transfer pipette to remove the DI water in the same way the filling solution was removed.
-
Repeat steps 2-4 two additional times.
-
Add a small amount of filling solution (See What is the correct storage (and filling) solution for an IntelliCAL pH probe?).
-
Shake the probe to rinse off the inside of probe of residual DI water with the filling solution.
-
Remove this small amount of filling solution with a transfer pipette.
-
Refill fully with fresh filling solution.