Question
What are the custom calibration restraints on the TL23 series benchtop turbidimeters?
Summary
Custom Calibration TL2300 TL2310 TL2350 TL2360 limitation
Answer
The constraint for custom calibration are:
1. The first point must be less than 0.1 and greater than 0
2. The second point less than or equal to 40 and greater than the first point
3. From the third point, Point[n] - Point[n-1] must be greater than or equal to 60
1. The first point must be less than 0.1 and greater than 0
2. The second point less than or equal to 40 and greater than the first point
3. From the third point, Point[n] - Point[n-1] must be greater than or equal to 60