What are the custom calibration restraints on the TL23 series benchtop turbidimeters?

Document ID

Document ID TE5289

Published Date

Published Date 12/20/2021
Print
Question
What are the custom calibration restraints on the TL23 series benchtop turbidimeters?
Summary
Custom Calibration TL2300 TL2310 TL2350 TL2360 limitation
Answer
The constraint for custom calibration are:

1. The first point must be less than 0.1 and greater than 0
2. The second point less than or equal to 40 and greater than the first point
3. From the third point, Point[n] - Point[n-1] must be greater than or equal to 60

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.