Question
What is the range, detection limit, and accuracy of the KF1000 Karl Fischer titrator?
Summary
Range, detection limit, and accuracy of the KF1000 Karl Fischer titrator.
Answer
Range and Detection limit:
The Volumetric Karl Fischer range on the KF1000 can be a little confusing, because it is in mg of water rather than in Percent moisture content or mg/L. The sample would need to have at least 1mg of water at its detection limit with an upper limit is 100mg of water. Percent moisture content range will then depend on sample amount used. The sample amount should not exceed 20g of sample. Range in Percent moisture content could be up 100% moisture content if the sample amount used is 0.1g or lower. For the lower limit, generally volumetric Karl Fischer is used for applications that contain 0.1% moisture content and higher, and for anything below 0.1% it would be recommended to do Coulometric titration. But the math shows that theoretically the instrument could go down as low as 0.005% {(1mg/20,000mg)*100= 0.005%.}, but generally you should speak with tech support so that they can ask a few questions if the application is expecting results below 0.1% to make sure it's going to be a good fit.
Accuracy and Precision:
There are many variables that impact accuracy and precision. Aside from normal things like lab technique and environment, there is also the accuracy of the sample amount measurement, if the sample amount is measured by pipette or balance, what reagents are used, the age of the reagents, the calibration of the titrant, and the condition and age of the electrode. This makes it impossible to determine. What would be recommended is to run standards.
The Volumetric Karl Fischer range on the KF1000 can be a little confusing, because it is in mg of water rather than in Percent moisture content or mg/L. The sample would need to have at least 1mg of water at its detection limit with an upper limit is 100mg of water. Percent moisture content range will then depend on sample amount used. The sample amount should not exceed 20g of sample. Range in Percent moisture content could be up 100% moisture content if the sample amount used is 0.1g or lower. For the lower limit, generally volumetric Karl Fischer is used for applications that contain 0.1% moisture content and higher, and for anything below 0.1% it would be recommended to do Coulometric titration. But the math shows that theoretically the instrument could go down as low as 0.005% {(1mg/20,000mg)*100= 0.005%.}, but generally you should speak with tech support so that they can ask a few questions if the application is expecting results below 0.1% to make sure it's going to be a good fit.
Accuracy and Precision:
There are many variables that impact accuracy and precision. Aside from normal things like lab technique and environment, there is also the accuracy of the sample amount measurement, if the sample amount is measured by pipette or balance, what reagents are used, the age of the reagents, the calibration of the titrant, and the condition and age of the electrode. This makes it impossible to determine. What would be recommended is to run standards.