What is the correct test to use if the sample water contains condensed phosphates?

Document ID

Document ID TE367

Published Date

Published Date 09/10/2020
Print
Question
What is the correct test to use if the sample water contains condensed phosphates?
Summary
Testing phosphorous when the water contains condensed phosphates
Answer
Phosphates which are condensed (meta-, pyro-, and polyphosphates) are inorganic and must be hydrolyzed to orthophosphate using acid and heat. The Hach method 8180 utilizes an acid persulfate digestion followed by PhosVer® 3 chemistry and should be used for condensed phosphates.

This procedure can be found at www.hach.com/WAH

Was this answer helpful?