When using a Hach color disc test kit, what is the purpose of the viewing tube without reagent in it?

Document ID

Document ID TE7112

Published Date

Published Date 09/11/2018
Print
Question
When using a Hach color disc test kit, what is the purpose of the viewing tube without reagent in it?
Summary
Viewing tube without reagent in color disc test kits
Answer
Hach color disc test kits require that two viewing tubes are filled with the unknown sample and then reagents are added only to one of the tubes. The viewing tube that does not contain reagent(s) is intended to be used as a blank. The blank is necessary so that sample qualities such as small amounts of turbidity and color do not interfere with the final result.