Question
Can an sc1000 display module be used with the sc1500 or SC4200c controllers?
Summary
sc1000 Display Module, sc1500 and SC4200c controller compatibility
Answer

The sc1500 and SC4200c controllers are not compatible with the sc1000 display module.  Any mobile device (tablet, phone, or PC) with an active internet connection is the display on Hach Claros Mobile Sensor Management (MSM).  If a local display is desired, it would be possible to use an analog output from the sc1500 into an sc200 or sc1000 leveraging an analog input card in the legacy controllers or a 3rd party 4-20 mA display device and configure these devices to display the assigned parameter measurement value.