Can the sc200 mA output card Product # 9334600 be used with the 5500sc analyzers?

Document ID

Document ID TE6916

Published Date

Published Date 03/02/2020
Print
Question
Can the sc200 mA output card Product # 9334600 be used with the 5500sc analyzers?
Summary
 sc200 mA output card on 5500sc analyzers.
Answer

The 4-20mA Output Expansion Module for SC200 Universal Controller Product # 9334600 mA output card will not function properly in a 5500sc analyzer. It will be recognized, and the card menu is present but there is not enough voltage to drive the outputs.
Product # 6792200 is the correct card for adding additional mA outputs to a 5500sc analyzer.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.