Can the Chloride Quantab® test strips be used to measure salt in cheese?

Document ID

Document ID TE6872

Version

Version 3.0

Status

Status Published

Published Date

Published Date 09/20/2018
Question
Can the Chloride Quantab® test strips be used to measure salt in cheese?
Summary
Chloride Quantab® for salt in cheese
Answer
The Chloride QuanTab® Test Strips, 30-600 mg/L (Product # 2744940) and the Chloride QuanTab® Test Strips, 300-600 mg/L (Product # 2751340) can be used to test salt in cheese by using the following instructions:

1. Add 90 mL of boiling water (RO/DI) to 10g of the finely grated sample in a 200 mL beaker.
2. Stir vigorously for 30 seconds, wait one minute, stir again for 30 seconds. Cool at 25°C.
3. Place filter paper, folded into a cone-shaped cup, point first into the beaker.
4. Place the lower end of the Quantab® into the filtrate (the solution which has seeped through the filter paper) being sure not to submerge titrator more than 1.0 inch.
5. 30 seconds after the moisture sensitive signal string at the top of the titrator turns dark blue or a light brown, record the Quantab® reading to the nearest 0.1 units on the titrator scale at the tip of the yellow-white peak.
6. Convert the Quantab® reading to percent sodium chloride (NaCl) or to ppm chloride (Cl-) using the calibration chart which is in included in the packaging. *Each lot of Quantab® has been individually calibrated by our QA department. Be sure to use the correct calibration chart, ie. the control number on the product being used must match the control number which is on the bottle.
7. Multiply the result by the dilution factor 10 to obtain the actual salt concentration in the sample.

Please see attached document for additional information and diagrams.
 
Attachments

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.