Do the QuanTab® test strips work in Acetonitrile?

Document ID

Document ID TE6866

Version

Version 3.0

Status

Status Published

Published Date

Published Date 09/20/2018
Question
Do the QuanTab® test strips work in Acetonitrile?
Summary
QuanTab® test strips in acetonitrile
Answer
The Chloride QuanTab® Test Strips, 30-600 mg/L (Product # 2744940) and Chloride QuanTab® Test Strips, 300-6000 mg/L (Product # 2751340) should react in Acetonitrile and other organic solvents as long as the wick does not get clogged. If the wick gets clogged, the sample could be extracted with water or acid and a dilution factor applied. Hach has not specifically tested in Acetonitrile, but similar solvents have been tested with success.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.