How much flow do I need to send to my BioTector Total Organic Carbon (TOC) Analyzer?

Document ID

Document ID TE6857

Published Date

Published Date 03/06/2019
Print
Question
How much flow do I need to send to my BioTector Total Organic Carbon (TOC) Analyzer?
Summary
BioTector TOC (B3500 and B7000) flow requirements.
Answer
Sample needs to be pulled from an ambient pressure source with the BioTector TOC analyzer. If using the Hach Sample Overflow Chamber, 19-BAS-031, we want to have 250-300 mL/min of flow. If a user creates their own sample vessel, we want that vessel to have enough flow to replace the contents 8 times per sample cycle to ensure proper analysis. Sample times are analyzer specific and can be checked within the menu.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.