Why is my 5500 AMC analyzer failing calibration?

Document ID

Document ID TE6855

Version

Version 3.0

Status

Status Published

Published Date

Published Date 03/27/2018
Question
Why is my 5500 AMC analyzer failing calibration?
Summary
5500 AMC analyzer calibration failure.
Answer
There are many reasons why the calibration might fail on the 5500 AMC analyzer. The most common problem is low absorption on the Cell 2 calibration results. This happens because the R3 is only used with Cell 2 during the calibration. Because the R3 is used so infrequently, that tube is prone to clogs or losing a prime. This is easily resolved or found by performing a delivery test on the R3 to Cell 2 via the Diag Menu. If your R3 is flowing and priming via this delivery test and does not help solve the calibration failure, please send us the data files from the unit for further troubleshooting.