What functionality is lost with the TU5300 sc or TU5400 sc when connected to an sc100 controller?

Document ID

Document ID TE6810

Version

Version 3.0

Status

Status Published

Published Date

Published Date 08/23/2018
Question
What functionality is lost with the TU5300 sc or TU5400 sc when connected to an sc100 controller?
Summary
Description of functionality reduction when a TU5300 sc or TU5400 sc is connected to an sc100 controller.
Answer
In general, the TU5300 sc and TU5400 sc can function when connected to an sc100. However, a few pieces of functionality may be lost or reduced. These are:
  • During menu operations (e.g. calibration / verification) you have to press a key twice. This is a known bug of the sc100 working with function codes.
  • Limited language set: English, German, Spanish, French, Italian, Netherland, Danish, Swedish, Chinese, Japanese, Portuguese
  • Limited display space (lines and characters per line). There will be loss of information in some screens (See pictures below; values are different because they were taken with instruments measuring different samples)
  • No capability for reminders (e.g. desiccant cartridge exchange)
  • No capability to download the service log for further instrument troubleshooting
  • Software updates are more complicated due to the sc100's requirement of a data transfer cable
Because of the loss of functionality above, and the potential of further functionality loss with future TU5 Series software updates, it is recommended to upgrade to a current Hach controller when a TU5300 sc or TU5400 sc will be connected.
 