Question
What RLU values are expected when conducting ATP standard calibrations using the LuminUltra PhotonMaster™?
Summary
Expected RLU values for the ATP standard calibration using the LuminUltra PhotonMaster™
Answer
When conducting ATP standard calibrations using the LuminUltra PhotonMaster™, different RLU values are expected depending on the type of Luminase used. The expected values are given below:
 
Type of Luminase Expected RLU range from a Fresh Bottle RLU value from an old bottle needing replacement
Luminase (Product # LU-3ML-FD) 15,000-25,000 5,000
LuminaseW (Product # LUW-3ML-FD) 2,000-3,000 500
LuminaseXL (Product # LUXL-3ML-FD) 150,000-250,000 50,000

If the RLU values are at or below the values listed in the third column of the table, the Luminase should be discarded and a new bottle should be rehydrated.